PURELAB Prima
The simple and dependable choice for primary grade water
The PURELAB Prima range from ELGA produces primary grade purified water from a potable source in a wide range of volumes for laboratory, medical and industrial applications. Using proven reverse osmosis (RO) technology, the PURELAB Prima is ideal for feeding ultra-pure water systems and for typical laboratory applications including glassware washing, a feed to autoclaves, and for use in environmental cabinets.



DESIGNED TO SUIT YOUR EXACT NEEDS
PURELAB Prima systems are compact in size and are available in a choice of flow rates to meet your specific needs. The smaller units provide flow rates of 7, 15 or 20 liters per hour whilst the larger units deliver 30, 60, 90 or 120 liters per hour. Sophisticated microprocessor monitoring helps ensure optimum performance at all times - and you can rely on easy, cost-effective and convenient operation, with user-friendly controls, informative status displays, and simple maintenance routines.



THE PURELAB PRIMA RANGE



Model 7 15 20 30 60 90 120
Water Quality Up to 98% rejection of inorganics; > 99% rejection of organics* & bacteria**; TOC typically < 0.1ppm
Flow Rates** (15 C) 7 L/hr 15 L/hr 20 L/hr 30 L/hr 60 L/hr 90 L/hr 120 L/hr
Purification Technologies Pre-treatment media to remove chlorine and gross particulate contamination,
with enhanced bacteriostatic operation Reverse osmosis
Typical Applications Feed to ultra-pure water systems, autoclaves, environmental cabinets and stills
Glassware washing
Steam generators
Hydroponics


* MW > 100 Dalton
** Subject to correct operating and maintenance procedures

TECHNICAL SPECIFICATIONS

Performance Specifications
Inorganics - min > 90% rejection > 90% rejection
Inorganics - typical up to 98% rejection up to 98% rejection
Organics* > 99% rejection > 99% rejection
TOC-typical < 0.1 ppm < 0.1 ppm
Product Specifications
Dimensions Height: 460mm
Width: 410mm
Depth: 270mm
Height: 740mm
Width: 570mm
Depth: 320mm
Weight (standard models) PURELAB Prima 7; 11.5kg
PURELAB Prima 15: 12.5kg
PURELAB Prima 20: 13.0kg
PURELAB Prima 30; 46.0kg
PURELAB Prima 60: 47.0kg
PURELAB Prima 90: 48.0kg
PURELAB Prima 120; 54.0kg
Operational Mode Displays
Power on
Process on
Standby
Auto-rinse
Power on
Process on
Standby
Auto-rinse
Disinfection
Safety Features Power fail-safe
Automatic tevel controls
Audio-visual alarm warnings
Auto-restart
Power fail-safe
Automatic level controls
Audio-visual alarm warnings
Auto-restart
Low feed pressure shut-off
System Monitoring RO permeate purity (uS/cm)
RO permeate temperaturę (C)
Reservoir level - % full
Consumable remaining life indicator
Feedwater quality (uS/cm)
RO permeate purity (uS/cm)
RO permeate temperature (°C)
% lon rejection
Consumable change reminder
Feedwater Specifications
Source Potable water Potable water
Maximum Fl 10 10
Maximum Conductivity 1400 uS/cm 1400 uS/cm
Free Chlorine <0,5 ppm <0,5 ppm
Temperature 1-35°C 1-35°C
Pressure Maximum
without boost pump
with boost pump
6 bar (90 psi)
bar (30 psi)
n/a
6 bar (90 psi)
Pressure Minimum
without boost pump
with boost pump
4 bar (60 psi)
flooded suction
n/a
1 bar (20 psi)