PURELAB Ultra

The PURELAB Ultra is the intelligent ultra-pure water purification system for your most critical applications. It incorporates many unique and innovative features that set it apart from other systems and guarantees water purity to 18.2MΩ-cm and beyond.



A model to meet your needs
The PURELAB Ultra range has 5 models specifically designed to deliver the ideal water quality for your bioscience or analytical requirements.



Life Science
ModelGeneticBioscience
Water Purity
Resistivity at 25°C18.2MΩ-cm
TOC (RO Pretreatment) 1 - 3 ppb3 - 10 ppb
Bacteria<0.1 CFU/ml1,2
Bacterial endotoxin <0.001 EU/ml <0.001 EU/ml
RNase and DNase Removed
Particles Ultrafilter
0.2µm point-of-use filter3
Purification Technologies Dual Labpure purification cartridges
Recirculation
Photo-oxidation (Genetic)
Ultrafiltration
Flow Rate Up to 2 liters/min
ApplicationsMolecular biology, microbiology, PCR, electrophoresis, DNA sequencing, mammalian cell culture, cell and culture media preparation, in-vitro fertilization, monoclonal antibody production


Analysis
ModelAnalyticIonic
Water Purity
Resistivity at 25°C18.2MΩ-cm
TOC (RO Pretreatment) 1 - 2 ppb 3-10 ppb
Bacteria<0.1 CFU/ml1,2
Particles 0.05µm ultra-microfilter
0.2µm point-of-use filter3
Purification Technologies Dual Labpure purification cartridges
Recirculation
Photo-oxidation
Ultra-microfiltration
Flow Rate Up to 2 liters/min
ApplicationsUltra-trace and trace inorganic and organic analysis, HPLC, GC-MS, ICP-MS, GF-AAS, TOC analysis, IC, electrochemistry


General Science
ModelScientific
Water Purity
Resistivity at
25°C
18.2MΩ-cm
TOC (RO Pretreatment) 3-10 ppb
Bacteria <1 CFU/ml1
Particles 0.2µm point-of-use-filter3
Purification Technologies Dual Labpure purification cartridges
Recirculation
Optional 0.2µm optional point-of-use filter
Dispense Rate Up to 2 liters/min
ApplicationsGeneral analysis, IC, AAS, ICP-ES, standard buffers

1With point-of-use filter 2<1 CFU/ml achievable without point-of-use filter 3Optional



PURELAB Ultra benefits at a glance
  • Unique PureSure System
    • Ensures purity beyond 18.2MΩ-cm
    • Extends the life of your purification cartridges
    • Provides advance warning of cartridge change so you are free to continue your research
  • Real time TOC monitoring provides total confidence in the organic quality of your ultra pure water
    • The system reports on Total Organic Carbon every 2 seconds
  • Full spectrum UV lamp for low microbial and organic specification
  • Efficient fast rinse ultrafiltration cartridge
    • Filters endotoxin, proteins and nucleases
    • Rinses up to 7 times faster than other endotoxin filters
    • Significantly reduces costs of sanitization
  • Automatic organics pre-rinse ensures optimum purity on start-up
  • Unique Volumetric Profile Dispense facility for easy recording, replication and speedy dispensing of multiple set volumes
  • Sensitive electronic dispense offers accurate flow control from single droplets up to 2 liters/min
  • New Labpure high capacity purification cartridges for longer life and lower running costs
  • PIN coded protection of critical operating parameters
  • Cartridge Identification technology provides validatable traceability and guarantees the quality of the cartridge
  • Automatic self calibration of electronic circuitry - takes just 4 seconds every time the system is in process mode
  • Unique disinfection regime ensures that all internal wetted components are sanitized
  • Market leading validation support manual available to meet GLP, FDA or other validation protocols